Health product recall

Progeny Vantage Panoramic X-Ray System with Cephalometric (2013-09-27)

Starting date:
September 27, 2013
Posting date:
October 28, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36391

Recalled products

A. Progeny Vantage Panoramic X-Ray System with Cephalometric 

Reason

This field action is being initiated to correct a recently identified real time controller (RTC)firmware error found in firmware versions 3.0, 3.1, 3.2, and 3.3. This error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. "failure to stop"). If this occurs while a patient is positioned under the column, there is the potential for injury, such as bruising, sprains, or resulting from quick movements to avoid the moving column.

Affected products

A. Progeny Vantage Panoramic X-Ray System with Cephalometric 

Lot or serial number

Serial Number: ZN44018

Model or catalog number
  • V5000C
Companies
Manufacturer
Midmark Corporation
675 Heathrow Drive
Lincolnshire
60069
Illinois
UNITED STATES