Progeny Vantage Panoramic X-Ray System with Cephalometric (2013-09-27)
- Starting date:
- September 27, 2013
- Posting date:
- October 28, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-36391
Recalled products
A. Progeny Vantage Panoramic X-Ray System with Cephalometric
Reason
This field action is being initiated to correct a recently identified real time controller (RTC)firmware error found in firmware versions 3.0, 3.1, 3.2, and 3.3. This error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. "failure to stop"). If this occurs while a patient is positioned under the column, there is the potential for injury, such as bruising, sprains, or resulting from quick movements to avoid the moving column.
Affected products
A. Progeny Vantage Panoramic X-Ray System with Cephalometric
Lot or serial number
Serial Number: ZN44018
Model or catalog number
- V5000C
Companies
- Manufacturer
-
Midmark Corporation
675 Heathrow Drive
Lincolnshire
60069
Illinois
UNITED STATES