This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Prestige SI Multi Purpose R/F Remote Controlled System

Starting date:
November 22, 2016
Posting date:
January 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-61640

Reason

GE Healthcare has recently become aware of a potential safety issue due to a Distar CRT monitor that has fallen off the monitor suspension related to product aging. A fall of a CRT monitor could result in a serious injury to a patient or operator.

Affected products

Prestige SI Multi Purpose R/F Remote Controlled System

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

S13001DP

Companies

Manufacturer

GE Healthcare

P.O. BOX 414

Milwaukee

53201

Wisconsin

UNITED STATES