This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Presource Packs by CARDINAL HEALTH (2015-06-23)
- Starting date:
- June 23, 2015
- Posting date:
- July 8, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54108
Recalled Product
- A.ROBOTIC TGH KIT
- STERILE OPEN HEART PACK
- STERILE BREAST AUGMENTATION PACK
Reason
Cardinal Health has determined the specimen pitchers (fg65-4357a the specimen pitcher, graduated, 1200 cc) contained in these presource kits may be brittle due to design changes. As a result of stacking the pitchers inside each other, breakage of the small internal ribs of the pitcher may occur.
Affected products
A.ROBOTIC TGH KIT
Lot or serial number
316358
Model or catalog number
PU30OTBTGE
Companies
- Manufacturer
-
Cardinal Health
1430 Waukegan Road
Waukegan
Illinois
UNITED STATES
B. STERILE OPEN HEART PACK
Lot or serial number
294544
314940
341064
356703
365963
674055
Model or catalog number
SCV30OHKGE
SCV30OHKGF
Companies
- Manufacturer
-
Cardinal Health
1430 Waukegan Road
Waukegan
Illinois
UNITED STATES
C. STERILE BREAST AUGMENTATION PACK
Lot or serial number
301289
319718
336952
374069
Model or catalog number
SBA30BROPH
Companies
- Manufacturer
-
Cardinal Health
1430 Waukegan Road
Waukegan
Illinois
UNITED STATES