Premier Solo Diamond – Large Inverted Cone
Brand(s)
Last updated
Summary
Product
Premier Solo Diamond – Large Inverted Cone
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Premier Solo Diamond – Large Inverted Cone | M0117646 M0120260 M0123294 | 807016C |
Issue
This recall is initiated due to the hardness not meeting the material specification for Premier Solo Diamond 807016C. Based on testing of the remaining blanks in stock, the defect in hardness is believed to affect less than 50% of the diamonds.
Recall start date: February 25, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Dental
Companies
Premier Dental Products Company
1710 Romano Drive, Plymouth Meeting, Pennsylvania, United States, 19462
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77105