Powerseal Curved Jaw Sealer & Divider Double Action
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Powerseal Curved Jaw Sealer & Divider Double Action |
More than 10 numbers, contact manufacturer. |
PS-0523CJDA |
Issue
Olympus conducted an investigation after receiving seven complaints globally regarding parts moving freely on the shaft of the device. Through the investigation, Olympus identified that certain lot numbers of Powerseal may have an out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring which affects the ability for the rotation knob to rotate the jaws of the device relative to the handle. The jaws may still be rotated by directly rotating the shaft of the device.
Recall start date: July 27, 2022
Additional information
Details
Gyrus Acmi, Inc
9600 Louisiana Ave. North, Brooklyn Park, Minnesota, United States, 55445
Get notified