Health product recall

POWER PROCESSOR SAMPLE PROCESSING SYSTEM (2015-10-05)

Starting date:
October 5, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55702

Affected Products

POWER PROCESSOR SAMPLE PROCESSING SYSTEM (2015-10-05)

Reason

Beckman Coulter has identified that the centrifuge module on the power processor sample processing system exceeds the published acoustic noise specification of 74 decibels. Beckman coulter implemented a new centrifuge motor on the power processor which included an additional cooling fan mounted directly on the motor shaft that spins when the centrifuge motor runs during centrifugation. As a result, centrifuges with the new motor have two cooling fans that are positioned very close to each other and spin in opposite directions causing air turbulence and acoustic noise.

Affected products

POWER PROCESSOR SAMPLE PROCESSING SYSTEM

Lot or serial number

90HP1301002

Model or catalog number

A11036