POWER PROCESSOR SAMPLE PROCESSING SYSTEM (2015-10-05)
- Starting date:
- October 5, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-55702
Affected Products
POWER PROCESSOR SAMPLE PROCESSING SYSTEM (2015-10-05)
Reason
Beckman Coulter has identified that the centrifuge module on the power processor sample processing system exceeds the published acoustic noise specification of 74 decibels. Beckman coulter implemented a new centrifuge motor on the power processor which included an additional cooling fan mounted directly on the motor shaft that spins when the centrifuge motor runs during centrifugation. As a result, centrifuges with the new motor have two cooling fans that are positioned very close to each other and spin in opposite directions causing air turbulence and acoustic noise.
Affected products
POWER PROCESSOR SAMPLE PROCESSING SYSTEM
Lot or serial number
90HP1301002
Model or catalog number
A11036
Companies
- Manufacturer
-
Beckman Coulter