Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59370

Affected products

  1. CONTROL BOARD FOR POWER PRO AMBULANCE COT
  2. POWER PRO IT AMBULANCE COT
  3. POWER PRO AMBULANCE COT

Reason

A few service reports alleged that "smoke" was coming from the foot end of the cot. This reportedly occurred when either the battery was placed into the cot or shortly after placing the battery into the cot. In addition to the alleged smoking, one of the reports had also noted that when the user attempted to lower the cot (prior to the alleged smoking) the device went into high speed retract and then a few moments later it began to smoke.

Affected products

A. CONTROL BOARD FOR POWER PRO AMBULANCE COT

Lot or serial number

Not applicable

Model or catalog number

6500002014

Companies
Manufacturer

Stryker Medical

3800 E Centre Avenue

Portage

49002

Michigan

UNITED STATES



B. POWER PRO IT AMBULANCE COT

Lot or serial number

150840805

Model or catalog number

6516000000

Companies
Manufacturer

Stryker Medical

3800 E Centre Avenue

Portage

49002

Michigan

UNITED STATES



C. POWER PRO AMBULANCE COT

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

6506000000

Companies
Manufacturer

Stryker Medical

3800 E Centre Avenue

Portage

49002

Michigan

UNITED STATES