Power-Pro 2
Brand(s)
Last updated
Summary
Product
Power-Pro 2
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Power-Pro 2 |
More than 10 numbers, contact manufacturer. |
650705550001 |
Issue
It has been identified that certain electrical cables may become damaged during ambulance loading/unloading. Damaged cables may result in loss of electric function, erratic/bouncy motion when raising or lowering the cot, or increased lowering speed. manual operation is not affected by this issue.
Recall start date: November 11, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies
Stryker Medical
3800 E Centre Avenue, Portage, Michigan, United States, 49002
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72210
Get notified
Receive emails about new and updated recall and safety alerts.