Health product recall

Power Express Centrifuge

Last updated

Summary

Product
Power Express Centrifuge
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Power Express Centrifuge More than 10 numbers, contact manufacturer. B36365 B36366

Issue

Customers have observed an issue where the metal part of the sample holder detaches from the plastic holder base when the robot arm grips a sample tube. If this occurs at a centrifuge, the metal part can inadvertently enter the centrifuge along with the sample tube. The centrifuge is unable to detect the presence of the detached metal part and begins to rotate. This can pose risks to both the equipment and the samples being processed. 

Recall start date: July 22, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75888

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe