Portrait HSWXB
Brand(s)
Last updated
Summary
Product
Portrait HSWXB
Issue
Medical devices - Performance
What to do
No impacted units have been sold in Canada. They are located at the importer's site.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Portrait HSWXB | SRW24120053SA SRW24120050SA | 5700219-11 |
Issue
GE Healthcare has become aware of a software issue that prevents "spo2 probe off" and "check resp patch" alarms initiated before or during reboot from being displayed after the reboot on certain Portrait™ Mobile Patient Monitor (hub) systems.
Recall start date: April 16, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
| Ge Healthcare Finland Oy |
| Kuortaneenkatu 2, Helsinki, Finland, 00510 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77359
Get notified
Receive emails about new and updated recall and safety alerts.