Portex Tracheal Intubation Stylet (2021-07-29)
- Starting date:
- July 29, 2021
- Posting date:
- August 19, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76255
Last updated:
Affected Products
Portex Tracheal Intubation Stylet
Reason
Smiths Medical became aware that one lot of Portex® Tracheal Intubation Stylets 335mm, Model Number 100/120/200, incorrectly contained Portex® Tracheal Intubation Stylets 365mm, Model Number 100/120/300. In the event that an incorrect tracheostomy stylet is not identified prior to use, a potential delay in placement may result.
Affected products
Portex Tracheal Intubation Stylet
Lot or serial number
4015424
Model or catalog number
100/120/200
Companies
- Manufacturer
-
Smiths Medical ASD, Inc.
6000 Nathan Lane North
Minneapolis
55442
Minnesota
UNITED STATES