Health product recall

Portable X-Ray Unit FDR Xair

Last updated

Summary

Product
Portable X-Ray Unit FDR Xair
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Portable X-Ray Unit FDR Xair

All lots.

XD2000 PX

Issue

The residual risk information of the XD2000 provided was not sufficient. There is the potential defectiveness for users to be exposed to unnecessary radiation by unintended handheld use. The users should understand the notification letter issued by FujiFilm and operate the XD2000 in accordance with the revised operation manual.

Recall Start Date: February 26, 2025
 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Fujifilm Corporation

26-30 Nishiazabu, Minato-Ku, Tokyo-To, Japan, 106-8620

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77064

Get notified

Receive emails about new and updated recall and safety alerts.