Health product recall

Polysorb Coated Braided Absorbable Single Stitch Reload For Use With The Endo Stitch And Sils Stitch Suturing Devices

Brand(s)
Last updated

Summary

Product
Polysorb Coated Braided Absorbable Single Stitch Reload For Use With The Endo Stitch And Sils Stitch Suturing Devices
Issue
Medical devices - Sterility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Polysorb Coated Braided Absorbable Single Stitch Reload For Use With The Endo Stitch And Sils Stitch Suturing Devices

J5H2924Y

170071

Issue

On Jan 22 2026, holes were found in both the breather pouch and the foil. During the investigation into the source of the issue, it was determined that the post seal machine was causing damage to the suture package.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Covidien Llc

15 Hampshire Street, Mansfield, Massachusetts, United States, 02048

Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Recall date
Identification number
RA-81924

Get notified

Receive emails about new and updated recall and safety alerts.