Health product recall

Poches À Sang Quadruple

Brand(s)
Last updated

Summary

Product
Poches À Sang Quadruple
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Poches À Sang Quadruple

11487575BM

LQT710X

Issue

A report on the batch 11487575bm reference lqt710x, 22 leaks on the plasma bags detected during processing. These leaks are located at the junction of top and bottom parts of the bags. The risk relates to a loss of integrity of plasma bag.

Recall start date: October 25, 2023
 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies
Macopharma
Rue Lorthiois, Mouvaux, France, 59420
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74564

Get notified

Receive emails about new and updated recall and safety alerts.