This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

POCHE A SANG QUADRUPLE (2015-12-08)

Starting date:
December 8, 2015
Posting date:
January 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56542

Affected Products

A. POCHE A SANG QUADRUPLE

Reason

Misassembled LCRD2 and MTL1 filters were reported to macopharma where the inlet and the outlet tubing were on the same side of the filter. This misassembly creates direct passage of blood from the collection bag to the transfer bag without passing through the filter

Affected products

A. POCHE A SANG QUADRUPLE

Lot or serial number

Not applicable

Model or catalog number

LQT7292LX

Companies
Manufacturer

Macopharma

Rue Lorthiois

Mouvaux

59420

FRANCE