POCHE A SANG QUADRUPLE (2015-12-08)
- Starting date:
- December 8, 2015
- Posting date:
- January 7, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56542
Affected Products
A. POCHE A SANG QUADRUPLE
Reason
Misassembled LCRD2 and MTL1 filters were reported to macopharma where the inlet and the outlet tubing were on the same side of the filter. This misassembly creates direct passage of blood from the collection bag to the transfer bag without passing through the filter
Affected products
A. POCHE A SANG QUADRUPLE
Lot or serial number
Not applicable
Model or catalog number
LQT7292LX
Companies
- Manufacturer
-
Macopharma
Rue Lorthiois
Mouvaux
59420
FRANCE