Health product recall

Poche A Sang Quadruple

Starting date:
November 24, 2016
Posting date:
January 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61616

Reason

Macopharma was informed of a number of fast filtrations on the NPT6280LA reference in use in France. This has caused an absence of effective filtration through the media of the LCRD2 Filters. This sporadic defect is linked to a partial absence of the weld in the media caused by poor positioning (fold or gap) of the last polyester layer in the mattress.

Affected products

Poche A Sang Quadruple

Lot or serial number

Not Applicable

Model or catalog number

LQT7292LX

Companies

Manufacturer

Macopharma

Rue Lorthiois

Mouvaux

59420

FRANCE