Plum™ Precision IV Pumps
Brand(s)
Last updated
Summary
Product
Plum™ Precision IV Pumps
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Plum Solo™ |
v1.0.3 |
40001-04-01 |
Plum Duo™ |
v1.2.3 |
40002-04-03 |
Issue
In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Icu Medical, Inc.
600 North Field Drive, Lake Forest, Illinois, United States, 60045
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82218
Get notified
Receive emails about new and updated recall and safety alerts.