Health product recall

Plum Infusion Systems

Last updated

Summary

Product
Plum Infusion Systems
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Plum A+ Infusion System All lots. 20792-XX-YY 12391-XX-YY
Plum A+ Hyperbaric Infusion System All lots. 11005
Plum 360 Infusion System All lots. 30010-XX-YY
Plum A+3 Infusion System All lots. 20678-XX-YY 20679-XX-YY 12618-XX-YY

Issue

ICU Medical has received reports of counterfeit CSB batteries being used with Plum infusion systems. While these counterfeit batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have not been tested or validated for use with Plum infusion systems. Preliminary reports suggest that these counterfeit batteries fail to hold their charge and the pump may display messages to replace batteries earlier than expected. Do not use spare parts that are not authorized by ICU Medical. Please ensure that any entity providing service or repair activities for your plum infusion systems uses only parts authorized or directly provided by ICU Medical.

Recall start date: October 21, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Health products - Medical devices - General and plastic surgery
Companies
ICU Medical, Inc.
600 North Field Drive, Lake Forest, Illinois, United States, 60045
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76494

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe