Plum Infusion Systems
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Plum A+ Infusion System |
All lots. |
20792-XX-YY |
Plum A+ Hyperbaric Infusion System |
All lots. |
11005 |
Plum 360 Infusion System |
All lots. |
30010-XX-YY |
Plum A+3 Infusion System |
All lots. |
20678-XX-YY |
Issue
ICU Medical has received reports of counterfeit CSB batteries being used with Plum infusion systems. While these counterfeit batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have not been tested or validated for use with Plum infusion systems. Preliminary reports suggest that these counterfeit batteries fail to hold their charge and the pump may display messages to replace batteries earlier than expected. Do not use spare parts that are not authorized by ICU Medical. Please ensure that any entity providing service or repair activities for your plum infusion systems uses only parts authorized or directly provided by ICU Medical.
Recall start date: October 21, 2024
Additional information
Details
| ICU Medical, Inc. |
| 600 North Field Drive, Lake Forest, Illinois, United States, 60045 |
Get notified
Receive emails about new and updated recall and safety alerts.