Health product recall

PLEX Elite 9000 (2018-10-22)

Starting date:
October 22, 2018
Posting date:
November 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68192

Affected products

PLEX Elite 9000

Reason

PLEX Elite 9000 power table up/down switch:  under certain fault conditions, the existing design may expose the  operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the PLEX  lite 9000 power table up/down switch. A hazardous voltage from an electrical short circuit could occur is a result of damaged insulation on one of the wires inside the enclosure for the power table up/down switch.

Affected products

PLEX Elite 9000

Lot or serial number
  • PE9000-0105
  • PE9000-0127
  • PE9000-0139
  • PE9000-0148
Model or catalog number

9000

Companies
Manufacturer
Carl Zeiss Meditec, Inc.
5160 Hacienda Drive
Dublin
94568
California
UNITED STATES