Health product recall

Plain Gut Absorbable Collagen Suture (2019-02-12)

Starting date:
February 12, 2019
Posting date:
March 1, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69184

Affected Products

Plain Gut Absorbable Collagen Suture

Reason

Routine monitoring of plain gut sutures not released to market exceeded endotoxin limits. As a precaution, suture lots manufactured from the same raw material lot are being recalled.

Affected products

Plain Gut Absorbable Collagen Suture

Lot or serial number

170927-52
180611-59
180924-51

Model or catalog number

810PG
822PG

Companies
Manufacturer
CP Medical Inc.
1775 Corporate Drive, Suite 150
Norcross
30093
Georgia
UNITED STATES