Health product recall

Philips Patient Information Center

Last updated

Summary

Product
Philips Patient Information Center
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Patient Information Center

Not applicable.

VERSION 4.0

Issue

During extended operation of the MX40 with the PIC iX, the DHCP (dynamic host configuration protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the pic ix, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.

Recall start date: May 12, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Philips Medizin Systeme Boeblingen GmbH

Hewlett-Packard Str. 2, Boeblingen, Germany, 71034

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73709