Health product recall

IntelliVue MX Patient Monitor System-IntelliBridge Module EC10

Last updated

Summary

Product
IntelliVue MX Patient Monitor System-IntelliBridge Module EC10
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

IntelliVue MX Patient Monitor System-IntelliBridge Module EC10

Not applicable.

865115

Issue

Potential safety issue concerning IntelliVue Patient Monitor (host monitor) with IntelliBridge EC10 Module and EC5 ID-Module interfacing with the Hamilton ventilator failing to alarm for "No Device Data" INOP.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Philips Medizin Systeme Boblingen Gmbh
Hewlett-Packard Strasse 2, Boblingen, Germany, 71034
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-76881

Get notified