Health product recall

Philips Achieva and Ingenia Systems

Last updated

Summary

Product
Philips Achieva and Ingenia Systems
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Achieva 3.0T MRI System - Main Unit

All lots.

ACHIEVA 3.0T

Ingenia 1.5T

All lots.

781315

Ingenia 1.5T

All lots.

1.5T

Ingenia 1.5T

All lots.

782115

Ingenia Elition X

All lots.

782107

Ingenia Elition X

All lots.

781358

MR 7700

All lots.

782153

Ingenia 3.0T CX

All lots.

781271

Ingenia 3.0T

All lots.

3.0T

Ingenia Ambition S

All lots.

782108

Ingenia Ambition S

All lots.

782139

Achieva XR System - Main Unit

All lots.

ACHIEVA XR

Ingenia Ambition X

All lots.

782109

Ingenia Ambition X

All lots.

781356

Ingenia Ambition X

All lots.

782138

Ingenia Elition X

All lots.

782136

Issue

Philips has identified an issue where component failures in the gradient coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

If a component failure occurs, the user may observe the following:

  • repeated scan abort error messages due to gradient amplifier fault detection
  • unusual noise from the system in the examination or technical room
  • burning smell
  • smoke and/or fire.

If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or death. This issue could also lead to property damage.

The scope includes MR systems with gradient coil types WA15, WA30S, WB30S, and AG30S.

Recall start date: July 7, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Philips Medical Systems Nederland B.V.

Veenpluis 6, Best, Netherlands, 5684 PC

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77750

Get notified