Health product recall

Pg 8528/Pg 8527/Pg 8535/ Pg 8536 Washer-Disinfectors

Starting date:
April 4, 2017
Posting date:
May 31, 2017
Type of communication:
Medical Device Recall
Subcategory:
Biologic/vaccine
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63392

Reason

An IT security vulnerability was discovered in a network penetration test carried out by an IT security expert. This only applies to machines which are actually connected to an in-house network. In the event of an attack on the internal network of a hospital, a laboratory or a surgery, there is a risk that (mostly encoded) data can be read out from the affected Miele washer-disinfectors.

Affected products

Pg 8528/Pg 8527/Pg 8535/ Pg 8536 Washer-Disinfectors

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

  • PG 8527
  • PG 8528
  • PG 8535
  • PG 8536

Companies

Manufacturer

Miele & Cie. KG

CARL-MIELE-STR. 29

Gutersloh

33332

GERMANY