This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Persona Distal Valgus Alignment Guide (2014-07-31)

Starting date:
July 31, 2014
Posting date:
August 8, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40981

Recalled products

  1. Persona Distal Valgus Alignment Guide

Reason

During the last 2 years, Zimmer has received 49 complaints related to binding of the Persona Distal Valgus Alignment Guide's locking knob. The root cause of the binding was determined to be excessive counter-clockwise rotation of the knob when unlocking the device. Current users are being provided with additional guidance on the locking/unlocking feature of the device.

Affected products

A. Persona Distal Valgus Alignment Guide

Lot or serial number
  • 62580270
  • 62624943
  • 62641120
  • 62662172
  • 62695918
  • 62749057
Model or catalog number
  • 42-5099-004-00
Companies
Manufacturer

Zimmer Inc.

1800 West Center Street

Warsaw

46580

Indiana

UNITED STATES