Permobil Powered Wheelchairs (2018-08-06)
- Starting date:
- August 6, 2018
- Posting date:
- August 17, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67446
Affected products
Permobil Powered Wheelchairs
Reason
Backrest support telescoping bars potentially fail in certain instances. The backrest support telescoping bars allow for the adjustment of the backrest height. Reports received consist of 1 or both back canes on the PS Backrest Frame Assembly, having broken while end-user is seated in the device. When this reported failure occurs, reports indicate the partial or complete detachment of the back frame from the seating which allows the end-user to lose positioning and fall backwards.
Affected products
Permobil Powered Wheelchairs
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
- PS C300
- PS C350
- PS C400
- PS C500
- PS M300
Companies
- Manufacturer
-
PERMOBIL INC