Health product recall

Percept PC DBS implantable neurostimulator (2021-03-16)

Starting date:
March 16, 2021
Posting date:
April 8, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75335



Last updated:
2021-04-08

Affected Products 

Percept PC DBS implantable neurostimulator

Reason

Notification to inform customers of a potential issue with the stimulation cycling setting when using the INS replacement feature on the A610 clinician programmer application version 2.0.4584, 2.0.4594, 2.0.4605. When replacing an INS and using the A610 "Replacement" workflow to transfer existing neurostimulator settings to the new neurostimulator, cycling may appear to be enabled but cycling is actually disabled. This issue can only occur when using the A610 "Replacement" workflow. For all current patients who have a replacement model B35200 Percept PC INS with cycling enabled, list of instructions have been provided. This issue is a software anomaly that affects how the A610 transfers cycling settings between INSS. Notification also informing customers of a potential software crash with the A610 clinician programmer application version 2.0.4584, 2.0.4594, 2.0.4605 when interrogating the Percept PC Implantable Neurostimulator (INS). Although the A610 Clinician programmer may crash, the INS continues to provide programmed therapy. Steps for mitigations provided in the notification.

Affected products

Percept PC DBS implantable neurostimulator

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

B35200

Companies
Manufacturer

MEDTRONIC CANADA ULC,

99 hereford street

Brampton

L6Y 0R3

Ontario

CANADA