Percept PC DBS implantable neurostimulator (2021-03-16)
- Starting date:
- March 16, 2021
- Posting date:
- April 8, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75335
Last updated:
2021-04-08
Affected Products
Percept PC DBS implantable neurostimulator
Reason
Notification to inform customers of a potential issue with the stimulation cycling setting when using the INS replacement feature on the A610 clinician programmer application version 2.0.4584, 2.0.4594, 2.0.4605. When replacing an INS and using the A610 "Replacement" workflow to transfer existing neurostimulator settings to the new neurostimulator, cycling may appear to be enabled but cycling is actually disabled. This issue can only occur when using the A610 "Replacement" workflow. For all current patients who have a replacement model B35200 Percept PC INS with cycling enabled, list of instructions have been provided. This issue is a software anomaly that affects how the A610 transfers cycling settings between INSS. Notification also informing customers of a potential software crash with the A610 clinician programmer application version 2.0.4584, 2.0.4594, 2.0.4605 when interrogating the Percept PC Implantable Neurostimulator (INS). Although the A610 Clinician programmer may crash, the INS continues to provide programmed therapy. Steps for mitigations provided in the notification.
Affected products
Percept PC DBS implantable neurostimulator
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
B35200