PEDIASAT OXIMETRY CATHETER SET (2021-07-13)
- Starting date:
- July 13, 2021
- Posting date:
- July 22, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-76075
Last updated: 2021-07-22
Affected Products
PEDIASAT OXIMETRY CATHETER SET
Reason
Edwards has received three (3) customer complaints reporting leaks during use. One complaint noted that after administering medication, the physician flushed a different lumen than the one where the medication was administered, and the patient’s vital signs responded to the medication. The other two complaints noted external leakage at the transition from the green hub (backform) to the tube. Edwards has confirmed a manufacturing defect leading to possible interlumen leakage.
Affected products
PEDIASAT OXIMETRY CATHETER SET
Lot or serial number
- 62728569
- 62889136
- 62903838
Model or catalog number
- XT3515SP
Companies
- Manufacturer
-
Edwards Lifesciences LLC
1 Edwards Way
Irvine
92614
California
UNITED STATES