This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
PC-1000 and PC-1000/LASER 1000 (2015-12-03)
- Starting date:
- December 3, 2015
- Posting date:
- January 7, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56534
Affected Products
- A. PC-1000/LASER 1000
- B. PC-1000
Reason
The recall relates to a potential problem with the jackscrew and nut assemblies on some panoramic X-ray model PC-1000 units. In some units, the jackscrew and nut assemblies may not have been adequately lubricated during assembly, which may result in excessive wear, and the assemblies may slip. Signs of excessive wear may include unusual noise or hesitation or interference with smooth operation while the unit is moving up or down.
Affected products
A. PC-1000/LASER 1000
Lot or serial number
6001 to 13885
Model or catalog number
PC-1000/LASER1000
Companies
- Manufacturer
-
Panoramic Rental Corporation
4321 Goshen Road
Fort Wayne
Indiana
UNITED STATES
B. PC-1000
Lot or serial number
6001 to 13885
Model or catalog number
PC-1000
Companies
- Manufacturer
-
Panoramic Rental Corporation
4321 Goshen Road
Fort Wayne
Indiana
UNITED STATES