Health product recall

Patient Table

Starting date:
March 16, 2017
Posting date:
May 31, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63318

Reason

Due to a potential manufacturing error, a rare possibility exists for a sudden drop in table height of the patient table. The maximum drop in height is 5cm and this error could potentially occur with or without user interaction. Therefore, it is necessary to inspect the table of distributed  systems to ensure it was manufactured in a manner that confirms safe operation.

Affected products

Patient Table

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

  • 10433519
  • 10433521
  • 10433522
  • 10433523
  • 10849722
  • 10849723

Companies

Manufacturer

Trumpf Medizin Systems GmbH T Co. Kg,

Karl-Zeiss-Str. 7-9,

Saalfeld,

07318