This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Patient Table
- Starting date:
- March 16, 2017
- Posting date:
- May 31, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-63318
Reason
Due to a potential manufacturing error, a rare possibility exists for a sudden drop in table height of the patient table. The maximum drop in height is 5cm and this error could potentially occur with or without user interaction. Therefore, it is necessary to inspect the table of distributed systems to ensure it was manufactured in a manner that confirms safe operation.
Affected products
Patient Table
Lot or serial number
More than 10 numbers, contact manufacturer
Model or catalog number
- 10433519
- 10433521
- 10433522
- 10433523
- 10849722
- 10849723
Companies
- Manufacturer
-
Trumpf Medizin Systems GmbH T Co. Kg,
Karl-Zeiss-Str. 7-9,
Saalfeld,
07318
GERMANY