Patient Information Center iX
Last updated
Summary
Product
Patient Information Center iX
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Patient Information Center iX | Not applicable. | VERSION 4.2 VERSION 4.0 VERSION 4.1 |
Issue
It has been discovered that when the user copies the customized event categories from one unit to another, the event catalog displays the customized event categories, but the changes are not actually applied. If a user reviews the events in the event monitor the user will see in the event details that the event category is not updated to reflect the customization. This issue is the result of cached changes to the event catalog not propagating in the database.
Recall Start Date: April 17, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Philips Medizin Systeme Boblingen Gmbh
Hewlett-Packard Strasse 2, Boblingen, Germany, 71034
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75444
Get notified
Receive notifications for new and updated recalls and alerts by category.