Patient Information Center iX
Summary
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Patient Information Center iX | Not applicable. | VERSION 4.0 VERSION 4.1 |
Issue
It was found that the care assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when ios device screens were locked with a passcode. The issue occurred on devices running IOS 15, 16 and 17 as well as multiple versions of iPhones. The care assist mobile application is not intended to replace the patient's primary alarming system. If users rely on the care assist application as their primary event notification alarming system, the failure of custom ringtones could result in a delay in treatment for patients experiencing a change in clinical condition.
Recall Start Date: March 6, 2024
Additional information
Details
Philips Medizin Systeme Boblingen Gmbh
Hewlett-Packard Strasse 2, Boblingen, Germany, 71034
Get notified
Receive emails about new and updated recall and safety alerts.