Health product recall

PAT CBL 4WIRE AHA SNAP HC (2020-04-23)



Starting date:
April 23, 2020
Posting date:
May 15, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73097


Last updated:

Affected Products

PAT CBL 4WIRE AHA SNAP HC

Reason

Internal testing has indicated that, in extremely rare cases, impacted Welch Allyn products may not meet the "Defibrillation Withstand" requirements of EN/IEC 60601-2-25 Medical Electrical Equipment. These are particular requirements for the Safety of Electrocardiographs; a standard the product claims to meet.

Affected products

PAT CBL 4WIRE AHA SNAP HC

Lot or serial number
  • Not applicable.
Model or catalog number
  • 9293-032-52
Companies
Manufacturer

Welch Allyn Inc.

4341 State Street Road

Skaneateles Falls

13153-0220

New York

UNITED STATES