Health product recall

Parapac Plus 310 Ventilators

Last updated

Summary

Product
Parapac Plus 310 Ventilators
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Parapac Plus 310 Ventilators

All lots.

P310NCAFREN
P310NCA
P310NUS

Issue

When a Parapac Plus™ ventilator is switched to the operating mode of 'ventilate', Parapac Plus™ ventilators may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This non‐cycling and continuous positive gas flow when in cycling mode, is a malfunction, not allowing the ventilator to properly function as designed.

Recall start date: February 26, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Smiths Medical Asd, Inc.
6000 Nathan Lane N., Minneapolis, Minnesota, United States, 55442
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75201

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe