Parapac Plus 310 Ventilators
Brand(s)
Last updated
Summary
Product
Parapac Plus 310 Ventilators
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Parapac Plus 310 Ventilators |
All lots. |
P310NCAFREN |
Issue
When a Parapac Plus™ ventilator is switched to the operating mode of 'ventilate', Parapac Plus™ ventilators may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This non‐cycling and continuous positive gas flow when in cycling mode, is a malfunction, not allowing the ventilator to properly function as designed.
Recall start date: February 26, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Smiths Medical Asd, Inc. |
6000 Nathan Lane N., Minneapolis, Minnesota, United States, 55442 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75201
Get notified
Receive notifications for new and updated recalls and alerts by category.