Health product recall

Paradigm and MiniMed 600 and 700 series pumps

Last updated

Summary

Product
Paradigm and MiniMed 600 and 700 series pumps
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalogue number

Minimed 630g

All lots.

MMT-1714

Paradigm 754 Insulin Pump

All lots.

MMT-754

Minimed 780g Kit

All lots.

MMT-1895

Minimed 670g

All lots.

MMT-1782

Paradigm 554 Insulin Pump

All lots.

MMT-554

Minimed 770g

All lots.

MMT-1891

Minimed 770g

All lots.

MMT-1881

Minimed 670g Kit

All lots.

MMT-1762

Minimed 780g Pump

All lots.

MMT-1885

Issue

As part of the testing for the Alternate Controller Enabled (ACE) device submission, the R&D team tested the Minimed 780g Insulin Pump at elevated and lowered height conditions relative to the infusion site per the engineering test report. The results of the testing showed that there is an unintended over or under delivery of insulin that occurs when the 780g pump is either elevated or lowered relative to the infusion site.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Medtronic Minimed

18000 Devonshire St., Northridge, California, United States, 91325

Published by
Health Canada
Audience
General public
Recall class
Type II
Recall date
Identification number
RA-81648

Get notified