Panocell-10 Reagent Red Blood Cells
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Panocell-10 (2-4% Suspension) Reagent Red Blood Cells |
21934A |
0003032 |
Panocell-10 Ficin Treated Reagent Red Blood Cells |
19919-E |
0002385 |
Issue
Immucor Inc. has indicated that this medical device recall is to communicate that some vials of indicated lot numbers exhibit an unexpected dark color (hemolysis) and darkening of the red blood cells. It has been confirmed that there is fungal growth in the vials demonstrating hemolysis and dark color. This issue is random across kits in these lots and is not present in all vials within the kit.
Recall start date: June 27, 2024
Additional information
Details
Immucor Inc. |
3130 Gateway Drive, Norcross, Georgia, United States, 30071 |