This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

P512 PRE-ANALYTICAL SYSTEM (2016-02-22)

Starting date:
February 22, 2016
Posting date:
April 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57996

Affected products 

A. P512 PRE-ANALYTICAL SYSTEM

Reason

It was observed that sample tubes may be decapped and then not correctly placed back in the rack tube transport. The sample material was spilled and could not be used to carry out the analysis.

Affected products

A. P512 PRE-ANALYTICAL SYSTEM

Lot or serial number

S/N: 4728845

Model or catalog number

5083435001

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY