This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

P-300 PORTABLE CEILING LIFT AND C-300 FIXED CEILING LIFT (2015-10-20)

Starting date:
October 20, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55772

Affected Products

  1. P-300 PORTABLE CEILING LIFT
  2. C-300 FIXED CEILING LIFT

Reason

Prism Medical Canada has been made aware of a number of incidents globally involving 300 Series Lifts in which the drive gear shaft has broken, eliminating the lifts’ ability to raise or lower a client. Of these incidents they have discovered very few cases in which the free fall protection system did not engage properly. If both of these incidents occur at the same time there is a risk for the person being supported by the lift to drop unexpectedly.

Affected products

A. P-300 PORTABLE CEILING LIFT

Lot or serial number

All lots

Model or catalog number
  • 303050
  • 303050R
  • 303051
Companies
Manufacturer

Prism Medical USA

45 Progress Parkway

Maryland Heights

63043

Maryland

UNITED STATES


B. C-300 FIXED CEILING LIFT

Lot or serial number

All lots

Model or catalog number
  • 330050
  • 330050R
  • 330053
  • 330053R
  • 330057
Companies
Manufacturer

Prism Medical USA

45 Progress Parkway

Maryland Heights

63043

Maryland

UNITED STATES