OXOID LEGIONELLA BCYE GROWTH SUPPLEMENT (2015-01-26)
- Starting date:
- January 26, 2015
- Posting date:
- March 5, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-52461
Recalled Products
OXOID LEGIONELLA BCYE GROWTH SUPPLEMENT
Reason
A technical investigation has found that specific Lots of Oxoid Legionella BCYE growth supplement (SR0110A) may not deliver adequate recovery of Legionella Pneumophila when used in conjunction with Oxoid Legionella medium (CM0655) and also with Oxoid Legionella selective supplement variants. Continued use of these Lots could result in false negative reporting.
Affected products
OXOID LEGIONELLA BCYE GROWTH SUPPLEMENT
Lot or serial number
1545171
1545172
Model or catalog number
SR0110A
Companies
- Manufacturer
-
Oxoid Limited
Wade Road
Basingstoke