This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

OXOID HAEMOLYSIN REAGENT (2015-05-07)

Starting date:
May 7, 2015
Posting date:
June 26, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Identification number:
RA-53920

Recalled Product

A.OXOID HAEMOLYSIN REAGENT

Reason

A technical investigation has indicated that the above lot may not react strongly with all strains of listeria Monocytogenes when used in the Microbact Listeria 12L kit (MB1128A) thus making interpretation of the result more difficult than with alternate batches. Continued use of this batch may lead to negative reporting.

Affected products

A.OXOID HAEMOLYSIN REAGENT

Lot or serial number

1330174

Model or catalog number

MB1249A

Companies
Manufacturer

Oxoid Limited

Wade Road

Basingstoke, Hampshire

RG24 8PW

UNITED KINGDOM