Outer-Grip Locking Fem Implant Impactor
Brand(s)
Last updated
Summary
Product
Outer-Grip Locking Fem Implant Impactor
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Outer-Grip Locking Fem Implant Impactor |
More than 10 numbers, contact manufacturer. |
74018902 |
Issue
Smith & Nephew, Inc. has initiated a field action to voluntarily remove several lots of outer-grip locking fem implant impactor due to a manufacturing error. There is an increased potential for the disassembly of the dowel pin and the impactor plate due to the reaming of the drive shaft pin holes.
Recall start date: March 7, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies
| Smith & Nephew, Inc. |
| 1450 Brooks Road, Memphis, Tennessee, United States, 38116 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77137
Get notified
Receive emails about new and updated recall and safety alerts.