Health product recall

OT 1000 SERIES ORTHOPEDIC SURGICAL TABLE (2021-01-22)



Starting date:
January 22, 2021
Posting date:
February 5, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74927


Last updated:

Affected Products 

OT 1000 SERIES ORTHOPEDIC SURGICAL TABLE

Reason

STERIS has identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.

Affected products

OT 1000 SERIES ORTHOPEDIC SURGICAL TABLE

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

OT 1100

OT 1200

Companies
Manufacturer

Steris Corporation

2720 Gunter Park East

Montgomery

36109

Alabama

UNITED STATES