Health product recall

ORTHO VISION Analyzer

Last updated

Summary

Product
ORTHO VISION Analyzer
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

ORTHO VISION Max Analyzer

More than 10 numbers, contact manufacturer.

6904576

ORTHO VISION Analyzer

More than 10 numbers, contact manufacturer.

6904577

Issue

During an internal investigation Quidelortho™ identified a potential for a false positive test result due to carryover of ORTHO Sera Anti-D (DVI) on vision analyzers.

Recall Start Date: November 8, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies

Ortho-Clinical Diagnostics Inc.

1001 Us Hwy 202, Raritan, New Jersey, United States, 08869

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76567

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe