Health product recall

Origio Sperm Wash

Brand(s)
Last updated

Summary

Product
Origio Sperm Wash
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Origio Sperm Wash

230922-018009

84055060D

Issue

During testing it was found that one of the batches, had an insufficient concentration of sodium pyruvate - only 0.002 mg/ml compared to a specified concentration of 0.046-0.137 mg/ml (1.5-4.3% of the specified amount).  Part of this formulation batch had also been used to create three lots of commercially released sperm wash medium.

Recall start date: January 22, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Obstetrics and gynaecology
Companies
Origio A/S
Knardrupvej 2, Malov, Denmark, 2760
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75188

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe