OR Table ORT300 (2014-09-25)
- Starting date:
- September 25, 2014
- Posting date:
- October 10, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-41741
Recalled Products
OR Table ORT300
Reason
One of the 3 pins to hold an alignment block to the floor of the MR suite was missing until during testing on the weekend the missing pin was pulled into the magnet, damaging the image quality phantom and headrest extension.
Affected products
OR Table ORT300
Lot or serial number
Part # 114093-000
Model or catalog number
114093-000
Companies
- Manufacturer
-
Imris Inc.
100-1370 Sony PLace
Winnipeg
R3T 1N5
Manitoba
CANADA