This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

OR Table ORT300 (2014-09-25)

Starting date:
September 25, 2014
Posting date:
October 10, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41741

Recalled Products

OR Table  ORT300

Reason

One of the 3 pins to hold an alignment block to the floor of the MR suite was missing until during testing on the weekend the missing pin was pulled into the magnet, damaging the image quality phantom and headrest extension.

Affected products

OR Table  ORT300

Lot or serial number

Part #  114093-000

Model or catalog number

114093-000

Companies
Manufacturer

Imris Inc.

100-1370 Sony PLace

Winnipeg

R3T 1N5

Manitoba

CANADA