Health product recall

Optima XR220amx X-Ray System

Last updated

Summary

Product
Optima XR220amx X-Ray System
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Optima XR220amx X-Ray System

SB2201600165WK
1035620WK2
1031494WK6
1027346WK4
00001027195WK5

5555000-5
5555000-6

Optima XR220amx X-Ray System

DF2401800480WK

5737545

Issue

GE Healthcare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the column safety mechanism can be missing.

Recall start date: December 4, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

GE Medical Systems, LLC

3000 North Grandview Blvd., Waukesha, Wisconsin, United States, 53188

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76714

Get notified

Receive emails about new and updated recall and safety alerts.