This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Optima X-Ray Systems (2013-12-05)
- Starting date:
- December 5, 2013
- Posting date:
- March 18, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-38245
Recalled Products
Optima X-Ray Systems
Reason
Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the mobile X-ray system are not properly displayed. If the user continues to attempt to select one of the exams when this issue has occurred, the system will update the screen and in doing so, potentially select a different exam/patient.
Affected products
A. Optima XR220AMX X-Ray
Lot or serial number
Not applicable
Model or catalog number
5555000-5
5555000-6
Companies
- Manufacturer
-
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Alabama
UNITED STATES
B. Optima XR200AMX X-Ray System - Digital
Lot or serial number
Not applicable
Model or catalog number
5421698
Companies
- Manufacturer
-
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Alabama
UNITED STATES