Optima Mr430s 1.5t - Main Unit
Brand(s)
Last updated
Summary
Product
Optima Mr430s 1.5t - Main Unit
Issue
Medical devices - Quality
What to do
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Optima Mr430s 1.5t - Main Unit | 403220MR1, 514697ONI, 905521ONI | OPTIMA MR430S 1.5T (5000-0002) |
Issue
GE healthcare (GEHC) MR systems could potentially have a cryogen ventilation system that does not meet the venting requirements. Failure to have proper venting could present a safety issue if the cryogen gas is vented into the MR room during a magnet quench, potentially depriving the room of oxygen.
Recall start date: Jan 4, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
| Ge Medical Systems, LLC |
|
301 Ballardvale St., Wilmington, Massachusetts United States, 01887 |
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63826