Optima CT660 and Revolution EVO (2017-06-01)
- Starting date:
- June 1, 2017
- Posting date:
- January 12, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65676
Affected products
- Optima CT660
- Revolution EVO
Reason
GE Healthcare has recently become aware of a potential safety issue that can result in a scan starting before the prescribed delay has completed. More specifically, if an exam is interrupted during the prep delay period using the emergency stop button, and the scan is then resumed, the delay period may be shorter than expected. If personnel are in the scan room at the time of the premature scan initiation, they can be exposed to scatter X-Ray radiation.
Affected products
A. Optima CT660
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
Optima CT660
Companies
- Manufacturer
-
GE Healthcare Japan Corporation
7-127 Asahigaoka, 4 Chome, Hinoshi
Tokyo
191-8503
JAPAN
B. Revolution EVO
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
Revolution EVO
Companies
- Manufacturer
-
GE Healthcare Japan Corporation
7-127 Asahigaoka, 4 Chome, Hinoshi
Tokyo
191-8503
JAPAN